myCROBE® solution – Instrument

Fully automated sample to result solution

CE IVD IVDR

Fully automated sample-to-result solution with a broad spectrum of diagnostic tests

The system covers the whole process from NA extraction, PCR setup (including automatic PCR plate sealing), PCR amplification to automatic result evaluation.

PCR tests currently supported in myCROBE can be found here.

Product Leaflet

Product Leaflet

; GP_MC_myCROBE_Solution_Leaflet_EN
pdf, [553.11 kB]
; GP_MC_Thrombotic_Mutations_Brochure 2023_DE
pdf, [865.35 kB]
; GP_MC_Thrombotic_Mutations_Brochure 2023_EN
pdf, [363.2 kB]
Choose all

System description

Capacity
1 – 36 samples / run
Up to 90 tests / run
Up to 108 samples and 270 tests / day
Time to result 4 hours
Hands-on time 30 minutes
User interface and control Integrated 10'' color touchscreen
Traceability QR codes labeling
Decontamination UV lamp
LIS connectivity Bidirectional connectivity (HL7, ASTM)
Remote access For diagnostics and system maintenance
Power supply 100 – 240 V, 50/60 Hz, 1200 W
Dimensions W x D x H 1130 x 750 x 900 mm
Weight 189 kg
Regulatory status CE IVD

Nucleic acids extraction

Capacity 1 – 36 samples / single extraction process
Extraction principle Extraction based on magnetic particles
Type of reagents used Single-use universal reaction cartridge
Sample volume 500 µl
Elution volume 100 µl
Extraction control myCROBE Universal Internal Control

PCR setup and analysis

PCR block capacity 96 reactions PCR plate
Parallel processing Up to 10 tests per sample
Quantification Universal set of 4 positive controls independent on the specific PCR
Automated analysis
No setup of analysis parameters necessary
Optional automatic sending the results to email
Free combinability Possibility to combine various examinations within 1 run / sample

Files are available only for registered users.

Instructions for Use & Advisory Notice

Instructions for Use are for informative purposes only

; GP_MC_Advisory_Notice_08_23
pdf, [775.8 kB]
; GP_MC_Instructions_for_Use_EN
pdf, [271.29 kB]

Declaration of Conformity

; GP_MC100_Declaration_of_Conformity_EN
pdf, [1.31 MB]

Manual

; GP_MC100_Manual_EN
pdf, [2.66 MB]

Videos