Low fluorescence
Possible Cause |
Solution |
Degraded MasterMix |
Store at -20 °C, thaw at 2 - 8 °C no longer than 30 min. Vortex and short spin before use. |
Unverified cycler |
Use verified cycler. List of verified cyclers here. |
Incorrect consumables |
Use recommended consumables. List of consumables in device manual. |
Clear PCR strips on BioRad CFX96 |
Use white PCR strips on BioRad CFX96 |
Incorrect transport medium |
Use TE buffer or Nuclease-free water. |
Air bubbles in PCR mix solution |
Spin PCR strip before PCR. |
Photosensitive MasterMix |
Minimize exposure of the MasterMix to light during handling to avoid loss of fluorescent signal intensity. |
No fluorescence
Possible cause |
Solution |
Incorrect PCR setup |
Contact support. Templates can be provided. |
Incorrect consumables |
Use recommended consumables. List of consumables in device manual. |
Incorrect transport medium |
Use TE buffer or Nuclease-free water. |
Photosensitive MasterMix |
Minimize exposure of the MasterMix to light during handling to avoid loss of fluorescent signal intensity. |
All